High Voltage Probe ACDC

We are engaged in manufacturing of wide assortment of High Voltage Probe ACDC, which are extensively used in various industries for applications that require you to take precise high-voltage measurements that need you to show accuracy. Additionally, our products are highly acclaimed for its precise measurement, high durability, sturdy construction and high performance. Our product can be availed in various specifications as per clients’ requirements. Moreover, we offer these instruments at virtually unmatched prices in the market.



High Voltage Probe

High Voltage Probe


Specifications :


Attenuate Ratio 40:00.00
Input Impedance Approx. 1000 M ohm.
Output Around 1.1 M ohm. The input impedance of external voltmeter
Impedance should be 10 mega ohm.
Safety Meet CAT II 40000 V
Max. Working Voltage DCV DC 40 KV

ACV Peak AC 40 KV or 28 KV rms ( depend which values is larger ).

DCV 1 KV to 20 KV - ± 1 %. 20 KV to 40 KV - ± 1.5 %.

ACV 1 to 28 KV rms, 50/60 Hz - ± 5 %.
Temp Coefficient Less than 200 ppm/.
Operating Temperature 0 to 50 ( 32 to 122 ).
Operating Humidity Less than 80% RH.
Cable Length 1 meter.


Operation :


Connect the plugs to the volts ( Hi ) & com ( Lo ) input terminals of your voltmeter ( or Multimeter ). Select the desired range of voltmeter ( Attention : Do not use auto ranging ). Whenever possible, turn the high voltage source off before making any connections. Connect the HV probe common lead ( alligator clip ) to a good earth ground or reliable chassis ground.


Safety Precaution & Warning :

  • This high voltage probe must be used by the person who are trained only. Do not work alone when working with high voltage circuits & environment.
  • For your own safety, inspect the probes for cracks & frayed or broken leads before each use. If any defects are noted, do not use the probes.
  • Hands, shoes, floor & work bench must be dry. Avoid making measurements under humid, damp or other environmental conditions that might affect the safety of measurement situation.
  • The ground connection must always be made before the probe tip comes into contact with the high voltage & must not be removed until after the probe tip has been removed from high voltage source.
  • Do not attempt to take measurement from sources where the chassis or return lead is not ground.
  • If possible, always turn the high voltage source off before connecting or disconnecting the probe.
  • Before turning the high voltage on, make sure that no part of your body is in contact with the device under test.
  • The probe body should be kept clean & free of any conductive contamination. Clean only the exterior probe body & cables. Use a soft cotton cloth lightly moistened with a mild solution of detergent & water. Do not allow any portion of the probe to be submerged at any time.








Contact Us
Global Instruments
Mr. Rahul Baid (Director)
No. 1688, Upper Ground Floor, Gali Jog Dhian Bhagirath Palace, Chandni Chowk
Delhi - 110 006, India

Contact via E-mail


Home  |  About Us  |  Our Product Range  |  Trade Offers (2)   |  Site Map  |  Contact Us

Member IndiaMART.com © Global Instruments. All Rights Reserved (Terms of Use)
Developed and Managed by IndiaMART InterMESH Limited

Business Enquiry Form


(* represents compulsory fields )

*Describe Your Requirements in Detail
Please Fill Your Contact Information
*Contact Person:
*Email:
Company:
Street Address:
*Country:
*Telephone:
Mobile:
Website:
*Enter the code: